Carence de validation pour ce fabricant de dispositifs médicaux...



Ce fabricant de dispositifs médicaux s'est fait épingler pour défaut de validation sur plusieurs systèmes impliqués dans la fabrication ou le contrôle des dispositifs médicaux :

2.    Software used as part of production and the quality system has not been validated for its intended use according to an established protocol, as required by 21 CFR 820.70(i). Specifically, software validation was not conducted for the following pieces of equipment:
 
Automated Production Equipment
Quality Inspection Equipment
(b)(4)
(b)(4)
(b)(4)
(b)(4)
(b)(4)
 
 
The corrective actions outlined in your firm’s responses appear adequate and will be verified during a during a follow-up FDA inspection.
 
Rappelons ce que précise le texte 820.70(i) sur ce sujet :
 
...When computers or automated data processing systems are used as part of production or the quality system, the manufacturer shall validate computer software for its intended use according to an established protocol. All software changes shall be validated before approval and issuance. These validation activities and results shall be documented. 
 
Plus d'information ici